1371 Prediction of Clinical Performance for Polymerization Devices with Different Optics-Design

Saturday, March 24, 2012: 9:45 a.m. - 11 a.m.
Presentation Type: Poster Session
J. GRAMANN, K. SCHEPKE-GERLACH, R. ZERGUINE, and A. SYREK, 3M ESPE AG, Seefeld, Germany
Objectives: The aim of this study was to predict the clinically relevant curing performance of polymerization devices having different optical concepts. In particular those concepts were:
1) Monochrome LED, reflector and glassfiber light guide with uniform diameter (3M ESPE, Paradigm™ Curing Light).

2) Monochrome LED, reflector and tapered glassfiber light guide (Kerr, DemiPlus).

3) Two differently colored LEDs (Polywave) and Fresnel lens without  light guide (Dentsply, Smartlite Max).

All devices are on a comparable power/intensity level.

Methods: Intensity of mentioned devices was measured with an Ocean Optics S2000 spectrometer and a 14mm optical sphere with 4mm aperture (according to ISO-4049 depth of cure (DOC) measurement setup). Intensity was measured at 0mm and 7mm distance between the tested device and the aperture.

In addition samples of Filtek™ Supreme Ultra (3M ESPE; shade A3B) were cured with the devices for 10s at 0mm and 7mm distance to measure DOC according to ISO-4049 (n=3; cylindrical sample with 4mm in diameter). The calculated DOC attenuation values were statistically analyzed by one-way ANOVA (p<0.05).

Results: The following values of intensity and DOC were found:

 

Intensity[mW/cm2]

DOC[mm]

Device

0mm

7mm

Attenuation [%]

0mm

7mm

Attenuation [%]

Paradigm™

2101±105

931±47

56±7

2.70±0.04

2.33±0.02

13.6±0.7

DemiPlus

2069±103

875±44

58±7

2.66±0.00

2.24±0.02

15.9±0.7

Smartlite Max

1870±94

525±26

72±6

2.27±0.01

1.65±0.01

27.2±0.3

With respect to the spectrometer’s resolution of 5%, Smartlight Max showed a higher attenuation in terms of intensity than Paradigm™ and DemiPlus.

Regarding DOC, Paradigm showed the significantly lowest attenuation whereas Smartlite Max showed the highest one.

Conclusions: Amongst the analyzed polymerization devices 3M ESPE Paradigm™ Curing Light showed the best curing performance at 0mm distance as well as at clinically relevant distance of 7mm between the device and the filling material. Also its intensity- and DOC-attenuation values are the best among the devices tested in this study.


Keywords: Composites, Curing lights and Polymerization